A new method only needs a tiny liquid sample to analyze metalloproteins. This breakthrough was achieved by a research team led by Associate Professor Eiji Ohmichi and Tsubasa Okamoto at the Kobe University Graduate School of Science. The findings were published on November 26 in Applied Physics Letters, and selected for Editor’s Pick as a particularly high-impact study.


Metalloproteins (also known as metal-binding proteins) play vital roles in our bodies for oxygen transport and storage, electron transport, oxidation and reduction. In many cases, the metal ions in these proteins are the active centers for these activities, so by identifying the exact state of these ions, we can understand the mechanisms behind their functions.

An experimental method called electron paramagnetic resonance (EPR) can be used to measure the state of electron ions in proteins. Effective EPR techniques require a certain amount of specimen volume for sensitive measurements. However, many metalloproteins are difficult to isolate and refine, so we can only obtain small samples.

Conventional EPR measurements detect the electromagnetic waves absorbed by metal ions. The notable feature of this study is the use of a trampoline-shaped device called a nanomembrane (figure 1(a)). In EPR the electron spin transitions to a high-energy state by absorbing electromagnetic waves, but at the same time the spin direction reverses, and the magnetic properties of the metal ions also change. Before the experiment the research team attached tiny magnets to the nanomembrane, so the changes in the force of attraction between the magnets and the metal ions are transformed into a force on the nanomembrane, and this EPR signal is detected. Since the nanomembrane is very thin (just 100 nm (=0.1 μm) we can sensitively measure small changes in force that accompany EPR absorption.

The solution specimen is placed in a solution cell directly above the membrane (figure 1(b)). The cell volume is just 50μL(=0.05 cc), and the team adds about 1-10μL(0.001-0.01 cc) of solution for measurement. In order to prevent the solution from evaporating, the cell is covered with a resin lid. In this method the thin and fragile nanomembrane is independent from the solution cell, making it easy to switch specimens (figure 1(c)).

Figure 1:

(a) Photo of a nanomembrane. The nanomembrane itself is very thin (100 nm) so it appears transparent to the naked eye. The item encircled in the center is a tiny magnet for EPR detection.
(b) Photo of the solution cell. The container is topped with a Teflon cover to prevent the solution sample evaporating.
(c) The experimental setup of the force-detected EPR spectroscopy. The solution cell containing the measurement sample is placed above the membrane with magnets attached. Changes in the nanomembrane are detected by the optical fiber below the membrane.

Figure 2: Results of the EPR measurements obtained in this study from a frozen solution sample

The top two diagrams are for hemin chloride, and the graph at the bottom is for myoglobin. The red line is a projected signal from a value simulator. The solution concentrations and sample volumes were 50 mM、2 μL for hemin chloride, and 8.8 mM、10 μL for myoglobin. Measurements took place at 4.2 K. By examining the two lines in the upper right graph we are able to determine the exact state of iron ions. The upper left graph shows the molecular structure of hemin chloride.

In order to evaluate the performance of this setup, the team carried out EPR measurement over a high-frequency (over 0.1 THz) for an iron-containing protein called myoglobin and its model complex hemin chloride (figure 2). The team succeeded in detecting EPR signals across a wide wave frequency (0.1-0.35 THz) for a 50 mM concentration, 2μL hemin chloride solution. They also observed a characteristic EPR signal for a 8.8 mM, 10μL specimen of myoglobin solution. A great advantage of this method is the ability to measure across a wide frequency range, making it applicable for metalloproteins with a variety of magnetic properties.

Professor Ohmichi comments: “This new method makes it possible to determine to a detailed level the state of the metal ions in a tiny amount of metalloprotein solution. We may be able to apply the method to metalloproteins that previously could not be measured. For example, in our metabolisms, a metalloprotein called peroxidase plays a crucial role by converting hydrogen peroxide into water, making it harmless, but the details of the mechanism for this reactive process are still unclear. The results from this study can potentially be applied as a leading analysis method to shed light on this sort of vital phenomenon.”

Journal information

Title
Force-detection of high-frequency electron paramagnetic resonance spectroscopy of microliter solution sample
DOI
10.1063/1.5055743
Authors
Tsubasa Okamoto, Hideyuki Takahashi, Eiji Ohmichi, Haruto Ishikawa, Yasuhisa Mizutani, and Hitoshi Ohta
Journal
Applied Physics Letters

Related links

Researchers